NEUBRESCOPE NBX-6166U

Long-Range Model Supporting Both Single- and Double-Ended Measurements

  • Brillouin measurement: PPP-BOTDA&BOTDR
  • Maximum measurement distance 100km
  • Maximum spatial resolution 2.5cm
  • Single-ended&Double-ended

The NBX-6166 has been upgraded to the NBX-6166U with a redesigned enclosure and improved spatial resolution from 10 cm to 2.5 cm.

Specification

Laser Wavelength 1550 ± 2 nm
Distance Range 100 m~100 km
Measurement Frequency Range 9~13 GHz
Range of strain measurements -30,000 to +40,000 με (-3% to +4%)
Measurement Frequency Scan Step 5, 10, 20, 50 MHz
Readout Resolution 5 cm (default), 2.5cm (minimum)
Sampling Points 600,000 (default), 3,000,000 (maximum)
Average Count Settings 2^5 ~ 2^23 times (incl, Hardware Average Count 2^16)
Function PPP-BOTDA
Pulse width 0.25 ns 0.5 ns 1 ns 2 ns 5 ns 10 ns -
Spatial resolution 2.5 cm 5 cm 10 cm 20 cm 50 cm 1 m -
Dynamic Range*1 0.5 dB 1 dB 2 dB 3.5 dB 6 dB 8 dB -
Measurement Range*2 0.5 km 1 km 5 km 10 km 20 km 25 km -
Function BOTDR
Pulse width 5 ns 10 ns 20 ns 50 ns 100 ns 200 ns 500 ns 1000 ns
Spatial resolution 0.5 m 1 m 2 m 5 m 10 m 20 m 50 m 100 m
Dynamic Range*1 3 dB 5 dB 8 dB 12 dB 15 dB 16 dB 22 dB 24 dB
Measurement Range*2 10 km 15 km 25 km 40 km 45 km 50 km 70 km 80 km
Strain Measurement Accuracy*3 PPP-BOTDA= 20με
BOTDR= 40με
Temperature Measurement Accuracy*3*4 PPP-BOTDA= 1.0 ℃
BOTDR= 2.0 ℃
Repeatability*3*4*5 PPP-BOTDA= 10 με / 0.5 ℃
BOTDR= 20 με / 1.0 ℃
Function COTDR
Pulse width 5 ns 10 ns 20 ns 50 ns 100 ns 200 ns 500 ns 1000 ns
Spatial resolution 0.5 m 1 m 2 m 5 m 10 m 20 m 50 m 100 m
Dynamic Range*1 10 dB 13 dB 16 dB 20 dB 23 dB 26 dB 30 dB 30 dB
Measurement Range*2 27 km 40 km 50 km 60 km 70 km 80 km 80 km 80 km
Applicable Sensing Fiber Single mode optical fiber
Connector Type FC-APC (factory default) / SC-APC (factory option)
Input/output Interface USB 2.0 x2, USB 3.0 x2, LAN x2, DisplayPort x1
Power Supply AC 100~240V, 50/60 Hz, 350 VA
Laser Class Class 1 (IEC60825-1 : 2001)
Dimensions / Weight approx. 450(W)×658(D)×295(H) mm / 35 kg
Operating Temperature 10~40 ℃, Humidity below 85 % (no dew condensation)
Storage Temperature 0~50 ℃
Place of Production Japan

Updated: 2026/04/22
*Specifications are subject to change without notice.
*1 Based on 2^16 average cycles.
*2 Based on average fiber loss of 0.3 dB/km using single mode fiber.
*3 Based on the measurement of strain-free, UV-coated fiber.
*4 Based on the measurement with strain-free, UV-coated fiber and in constant temperature environment.
*5 Maximum standard deviation of measurement value in 5 consecutive measurements for 100 consecutive points.
*1-*5 are all based on a frequency scan step of 5 MHz with Pre-Pump Adjustment on.