NEUBRESCOPE NBX-5101U

Long-Range Single-Ended Measurement Model

  • BOTDR (Brillouin) Measurement
  • Max. Measurement Distance: 100km
  • Highest Spatial Resolution: 50cm
  • Single-ended

The NBX-5101 has been upgraded to the NBX-5101U with a redesigned enclosure of a different size.

Specification

Laser Wavelength 1550 ± 2nm
Distance Range 100m ~ 100km
Measurement Frequency Range 9 ~ 13GHz
Range of strain measurements -30,000 to +40,000με (-3% to +4%)
Temperature Measurement Range -150 °C ~ +600 °C (fiber dependent)
Measurement Frequency Scan Step 5, 10, 20, 50 MHz
Readout Resolution 5 cm (default), 2.5cm (minimum)
Sampling Points 600,000 (default), 3,000,000 (maximum)
Average Count Settings 2^8 to 2^24 (incl. 2^16 hardware counting)
Pulse width 5ns 10ns 20ns 50ns 100ns 200ns 500ns 1000ns
Spatial resolution 0.5m 1m 2m 5m 10m 20m 50m 100m
Function BOTDR
BOTDR Dynamic Range (dB)*1 3dB 5dB 8dB 12dB 15dB 16dB 22dB 24dB
BOTDR Measurement Distance (km)*2 10km 15km 25km 40km 45km 50km 70km 80km
Function COTDR
COTDR Dynamic Range (dB)*1 10dB 13dB 16dB 20dB 23dB 26dB 30dB 30dB
COTDR Measurement Distance (km)*2 27km 40km 50km 60km 70km 80km 80km 80km
Strain Accuracy (1σ)*3 20με 15με
Temperature Accuracy (1σ)*3*4 1.0 ℃ 0.5 ℃
Repeatability*3*4*5 10 με / 0.5℃ 7.5 με / 0.25℃
Fiber Channel 2
Suitable fiber Single mode optical fiber
Fiber Connector FC/APC (factory default) / SC-APC (factory option)
Input/Output Interfaces USB 2.0 x2, USB 3.0 x2, LAN x2, displayport x1
Power Supply

AC 100~240V, 50/60 Hz, 350 VA

Laser Class Class 1 (IEC60825-1 : 2001)
Dimensions / Weight approx. 450(W)×658(D)×295(H) mm / 35 kg
Operating Temperature 10~40 ℃, Humidity below 85 % (no dew condensation)
Storage Temperature 0~50 ℃
Place of Production Japan
Updated: 2026/04/22
*1 Based on 215 average cycles.
*2 Based on average fiber loss of 0.3 dB/km using single mode fiber.
*3 Based on the measurement of strain-free, UV-coated fiber.
*4 Based on the measurement of strain-free, UV-coated fiber and in constant temperature environment.
*5 Maximum standard deviation of measurement value in 5 consecutive measurements for 100 consecutive points.