You are at... HOME > Products > NEUBRESCOPE >Lineup> NBX-N9000 Series (Specs)
NEUBRESCOPETM
NBX-N9000 Series Specifications / Catalog
Function |
ERA-BOTDA |
OTDR |
|||||||
NBX-N9010 |
1650 ± 1 nm |
1650 ± 1 nm |
|||||||
NBX-N9020 |
1625 ± 1 nm |
1625 ± 1 nm |
|||||||
Measurement Type |
Pulsed probe pump Brillouin Analysis |
OTDR |
|||||||
Main Features |
Link Loss, Branch Loss |
Loss |
|||||||
Measurement Frequency Range |
8 ~ 12 GHz |
- |
|||||||
Range of Frequency Swept |
100MHz ~ 600MHz |
- |
|||||||
Frequency-Swept Speed of Pump Pulse |
0.1MHz/ns ~ 2MHz/ns |
- |
|||||||
Distance Range |
1km ~ 20km( 1km/step) |
||||||||
Sampling Points |
100,000 (default) |
||||||||
Sampling Interval |
0.5m, 1m, 2m, 5m, 10m, 20m |
||||||||
Resolution of Reflective Probe |
1m, 2m, 5m, 10m, 20m |
||||||||
Average Count Settings |
25 ~ 223 times |
||||||||
Function |
ERA-BOTDA |
OTDR |
|||||||
Pump Pulse Width (ns) |
50 |
10 |
|||||||
Spatial Resolution (m) *1 |
2 |
5 |
10 |
20 |
40 |
1 |
2 |
5 |
10 |
Dynamic Range of |
14 |
16 |
20 |
22 |
24 |
- |
|||
Dynamic Range of OTDR
(Back Scattering, dB) *3 |
- |
5 |
8 |
15 |
18 |
||||
Measurement Accuracy *4 |
0.5dB |
1dB |
|||||||
Repeatability *4 |
0.2dB |
0.5dB |
|||||||
Measurement Time *5 |
3 minutes |
30 seconds |
|||||||
Applicable Fiber |
Single-mode Fiber |
||||||||
Connector Type |
SC-UPC (Factory option) |
||||||||
Interface | USB 3.0 x4, LAN (RJ-45) x2, RGB x1 | ||||||||
Power Supply |
AC100 ~ 240V, 50/60Hz, 250VA |
||||||||
Laser Safety Class |
Class 1 (IEC60825-1 : 2001) |
||||||||
Dimensions / Weight |
approx. 456 (W) × 485 (D) × 286 (H) mm / 30 kg |
||||||||
Operating Temperature |
10~40 °C, Humidity below 85% (no dew condensation) |
||||||||
Storage Temperature |
0 ~ 50 °C |
||||||||
Place of Production |
Japan |
*1. Based on calculation result of Frequency-Swept Speed of Pump Pulse.
*2. Based on 2^16 average cycles. The Total Loss = Link Loss + Return Loss of FTTx (or OTDR) Reflector.
*3. Based on 2^16 average cycles. The Total Loss = Link Loss
*4. The maximum standard deviation of measurement value in 5 consecutive measurements for 100 consecutive points.
*5. The settings of 10k m distance range, 2^14 count settings, 100 sampling points not excluding the time for Pulse Adjustment.
(2018.05.21) |
Neubrex Co., Ltd. Sakaemachidori 1-1-24, Chuo-ku, Kobe, Hyogo, 650-0023 JAPAN [map] Copyright © 2007 Neubrex Co., Ltd. Tel:+81-78-335-3510/Fax:+81-78-335-3515 E-Mail: info@neubrex.jp
|