You are at... HOME > Products > NEUBRESCOPE >Linup> NBX-6026 (Specs)
NEUBRESCOPE™ (Patent pending)
NBX-6026 Specifications ・ / Catalog
laser wavelength |
1550 ± 2nm |
||||
measuring object |
strain distribution,temperature distribution,center frequency distribution,loss distribution,BGS width distribution,SBS distribution,BGS |
||||
distance range |
50m, 100m, 250m, 500m, 1km, 2.5km, 5km, 10km, 25km |
||||
measurement frequency range |
9 to 13 GHz |
||||
range of strain measurement |
-30,000 to +40,000με (-3% to 4%) |
||||
measurement frequency scan step |
1,2,5,10,20,50 MHz |
||||
readout resolution |
5cm (default), 1cm(minimum) |
||||
average count settings |
2^5 to 2^23 (inc. Hardware Average Count 2^16) |
||||
pulse width |
0.2 ns |
0.5 ns |
1 ns |
2 ns |
|
spatial resolution |
2 cm |
5 cm |
10 cm |
20 cm |
|
dynamic range *1 |
0.5 dB |
1 dB |
1.5 dB |
3 dB |
|
max. measurement distance *2 |
0.5 km |
1 km |
2 km |
5 km |
|
optical budget *1*6 |
2 dB |
3 dB |
5 dB |
7 dB |
|
Measurement accuracy *3*4 |
15 με / 0.75 °C |
7.5 με / 0.75 °C |
|||
Repeatability *3*4*5 |
10 με / 0.5 °C |
7.5 με / 0.35 °C |
|||
measurement Time*7 |
10 seconds (minimum,
Readout Resolution: 5cm) |
||||
siginal terminal ・ |
signal fiber |
SM fiber・ |
|||
fiber connector |
FC-APC / SC-APC (factory option) |
||||
suitable fiber |
SM fiber |
||||
Input/Output/Control Interface |
USB 2.0 x4, LAN x2, RGB x1 |
||||
power supply |
AC100 to 240V, 50/60Hz, 250VA |
||||
laser class |
CLASS 1 (IEC60825-1: 2001) |
||||
dimensions / weight |
approx. 456(w) x 485(D) x 286 (H)mm / 30kg |
||||
operating temperature |
10°C to 35°C / humidity below 85% (no dew condensation) |
||||
storage temperature |
0 to 50°C |
||||
place of production |
Japan |
||||
(2018/06/18) |
|||||
*1 Based on 2^15 average cycles by progressive measurement mode. |
|||||
*2 Based on average fiber loss of 0.3 dB/km using SM fiber |
|||||
*3 Based on the measurement of strain-free UV-coated fiber |
|||||
*4 Based on the measurement of strain-free UV-coated fiber and in constant temperature environment |
|||||
*5 The maximum standard deviation of measurement value in 5 consecutive measurements for 100 consecutive points |
|||||
*6 Within the allowable range adjusted by the optical power excluding the case of nonlinear phenomena |
|||||
*7 The setting of 50m range, 2^14 count settings, 41scan steps excluding the time for Pulse Adjustment |
|||||
*1-*6 are all based on a frequency scan step of 5 MHz and with Pulse Adjustment and Auto Frequency Adjustment on. |