You are at... HOME > Products > NEUBRESCOPE >Linup> NBX-60555(PM) (Specs)
NEUBRESCOPE? (Patent pending)
NBX-6055PM Specifications ・
| laser wavelength | 1550ア2 nm | |||||
| polarizaion extinction ratio | more than 20 dB | |||||
| measuring object | Strain, Temperature, SBS, BGS center
frequency, BLS center frequency, Loss |
|||||
| distance range | 50m, 100m, 250m, 500m, 1km, 2.5km, 5km, 10km, 25km | |||||
| measurement frequency range | 9 ? 13 GHz | |||||
| range of strain measurement | -30,000 to +40,000με (-3% to 4%) | |||||
| measurement frequency scan step | 1,2,5,10,20,50 MHz | |||||
| readout resolution | 5cm (default), 1cm(minimum) | |||||
| average count settings | 2^5 ? 2^23 (inc. Hardware Average Count 2^16) | |||||
| pulse width | 0.5 ns | 1 ns | 2 ns | 5 ns | 10 ns | |
| spatial resolution | 5 cm | 10 cm | 20 cm | 50 cm | 100 cm | |
| dynamic range *1 | 1 dB | 2 dB | 3 dB | 5 dB | 7 dB | |
| max. measurement distance *2 | approx. 1 km |
approx. 5 km |
approx. 10 km |
approx. 15 km |
approx. 25 km |
|
| optical budget *1*6 | 3 dB | 5 dB | 7 dB | 8 dB | 10 dB | |
| Measurement accuracy *3*4 | 10 με / 0.5 ℃ |
5 με / 0.2 ℃ | ||||
| Repeatability *3*4*5 | 7.5 με
/ 0.35 ℃ |
3.5 με / 0.15 ℃ | ||||
| measurement Time*7 | 5 seconds (minimum) | |||||
| siginal terminal ・ | signal fiber | PANDA fiber・ | ||||
| fiber connector | FC-PC / SC-PC (factory option) | |||||
| Polarization Plane | Perpendicular (Slow axis, -Y) | |||||
| suitable fiber | PANDA fiber | |||||
| power supply | AC100 ?240V 50/60Hz 250VA | |||||
| laser class | CLASS 1 (IEC60825-1: 2001) | |||||
| dimensions / weight | approx. 456(w) x 485(D) x 286 (H)mm / 30kg | |||||
| operating temperature | 10 ℃? 35 ℃ / humidity below 85% (no dew condensation) | |||||
| storage temperature | 0 ?50℃ | |||||
| place of production | Japan | |||||
| (2012/03/30) | ||||||
| *1 Based on 2^15 average cycles by progressive measurement mode. | ||||||
| *2 Based on average fiber loss of 0.3 dB/km using PANDA fiber | ||||||
| *3 Based on the measurement of strain-free PANDA fiber | ||||||
| *4 Based on the measurement of strain-free PANDA fiber and in constant temperature environment | ||||||
| *5 The maximum standard deviation of measurement value in 5 consecutive measurements for 100 consecutive points | ||||||
| *6 Within the allowable range adjusted by the optical power excluding the case of nonlinear phenomena | ||||||
| *7 The setting of 50m range, 2^14 count settings, 41scan steps excluding the time for Pulse Adjustment | ||||||
| *1-*6 are all based on a frequency scan step of 5 MHz and with Pulse Adjustment and Auto Frequency Adjustment on. | ||||||