laser wavelength 1550±2 nm
measuring object strain distribution,temperature distribution,center frequency distribution,loss distribution,BGS width distribution,SBS distribution,BGS
distance range 50m, 100m, 250m, 500m, 1km, 2.5km, 5km, 10km, 25km
measurement frequency range 9 ∼ 13 GHz
range of strain measurement -30,000 to +40,000με (-3% to 4%)
measurement frequency scan step 1,2,5,10,20,50 MHz
readout resolution 5cm (min)
average count settings 2^5 ∼ 2^23
pulse width 0.2 ns 0.5 ns 1 ns 2 ns
spatial resolution 2 cm 5 cm 10 cm 20 cm
dynamic range *1 0.5 dB 1 dB 1.5 dB 3 dB
max. measurement distance *2 100 m 1 km 2 km 5 km
optical budget *6 1 dB 3 dB 5 dB 7 dB
strain measurement accuracy*3 ±25 με
temperature measurement accuracy *4 ±1 ℃
repeatability*3*4*5 ±10 με / ±0.5 ℃
measurement speed*7 200 seconds(min)
siginal terminal signal fiber SM fiber
fiber connector FC-APC / SC-APC (factory option)
suitable fiber SM fiber
power supply AC100 ∼240V 50/60Hz 250VA
laser class CLASS 1 (IEC60825-1: 2001)
dimensions rack type
operating temperature 10 ℃∼ 35 ℃ / humidity below 85% (no dew condensation)
storage temperature 0 ∼50℃
place of production Japan
(2008/04/02)
*1 Based on 2^15 average cycles
*2 Based on average fiber loss of 0.3 dB/km using SM fiber
*3 Based on the measurement of strain-free SM fiber
*4 Based on the measurement of strain-free SM fiber and in constant temperature environment
*5 The deviation range of measurement value for 5 consecutive measurements for 100 consecutive points
*6 Within the allowable range being adjusted by the optical power, except the case of nonlinear phenomena
*7 Within the setting of 50m range, 2^13count settings, 41scan steps except the time of Pre-Pump Adjustment
*1-*5 are based on a frequency san step of 5 MHz and with Pre-Pump Adjustment and Auto Frequency Adjustment on.